Balance connection, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 48

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 48
Balance
connection
Connect the balance to the Titration Manager using the cable corre-
sponding to the type of balance used, see table below:
Table 2: Balance types
Cables can be made to your desired specifications. Please contact our
sales representative if you require information regarding the type of
cable to use with your balance.
Format of messages accepted by the Titration Manager:
•
M
essages end: LF or CR, LF+CR,
•
Units, g, mg, kg,
•
The character “g” of the unit is required to mark the end of the data
recognised by the instrument,
•
Spaces are ignored,
•
The number of characters situated between the first character in the mes-
sage and the “g” in the unit must not be more than 30, (spaces not
included),
•
The number of characters received after the “g” and before the end of the
message LF, is insignificant,
•
The numeric data of the weight must be transmitted before the unit.
RS232C parameters (to set on the balance):
2400 baud, even parity, 7 data bits and 1 stop bit
Example of a message transferred from a balance, type Mettler
The units selected in the Titration Manager must be identical to the
units used by the balance.
Balance
Type
Cable
Length
Mettler
PE + option 016
AE + option 012 or 011
A95Z201
A95Z205
1 m
2 m
AM, PM, SM, AT, MT,
UMT, PJ, AJ
A95Z206
1 m
Sartorius
MasterPro, QC, MP8-4
BP110S, CT224S,
LE244S-OCE
A95Z202
1 m
A95Z204
2 m
Precisa
All models (except XB220)
A95Z207
1 m
XB220
A95Z208
1 m
S
I
1
2
.
3
4
5
6
g
CR
LF
data received: 12.3456 g