Electrode connection, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 86

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 86
Electrode
connection
1.
In the Main window, press 2 Select method and select the method
to be run. The electrodes to be connected are listed with their
addresses.
2.
Install electrodes on the Titration Manager (TIM), or sample
changer (SAC80 or SAC850).
3.
Connect electrodes to the rear panel socket of the Titration
Manager (TIM). See figure and table below. Example: pHC2401-8
to E1 socket (address TIM/E1).
Figure 15: Electrode sockets
Table 3: Connecting electrodes
When installing/checking an electrode system, do not forget to
connect all the electrodes to the Titration Manager and disconnect all
the electrodes that are not used by the method or sequence.
In particular, if you are about to run a pH (or potential at zero current)
measurement, check that there is no double metal electrode con-
nected to the Pt-Pt instrument socket before running the method (dis-
connect this electrode, if any).
P
t
-P
t
R
ef
E1
E2
T
emp
GND
Socket
Electrode
REF
Single reference
TEMP
Temperature
GND
Single metal for cell grounding only
Pt-Pt
Double metal
E1
Others