Glp-archives menu, Ground conflict, Help – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 104: High (result indicator), See "glp-archives menu, Titra lab

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 104
GLP-Archives
menu
The GLP-Archives (Good Laboratory Practice) command is available
in each window:
To access:
•
Sample results: enter Main window and press 4
•
Reagent calibration results: enter Reagent window and press 4
•
Electrode calibration results: enter Electrode window and press 3
Figure 20: GLP-Archives for sample results
Storage capacity:
Last 60 sample results.
Last calibration of each reagent.
Last calibration of each electrode.
When the GLP-Archives is full and a new result arrives, the
oldest result stored will be the first one removed.
Ground conflict
Ground conflict: External grounding defined in Setup/Configura-
tion and a metal electrode.
An external grounding is defined for the titration system cell in the
Configuration screen and a metal electrode is used by the method.
When a metal electrode is used, select TIM cell external
Gnd = No in the Configuration menu.
Help
Refer to "Check command", page 59
High (result
indicator)
Refer to "Result indicators", page 157
.
The RIGHT and LEFT arrow keys allow you to
move from one screen to the other.
1/4 means the first results out of 4 results saved
for the selected method.