Titre Enter/
Calibrate

Enter

If you already know the reagent concentration and do not wish to per-
form a calibration.

Calibrate

If you wish to perform a calibration. The Titration Manager will prompt
you to enter the calibration parameters.

It is not necessary to calibrate an addition reagent. It is there-
fore recommended to select Titre = Enter so that the calibration
parameters are not displayed.

Enter in:

Edit reagent menu.

Tray missing
(SAC error)

On starting a sequence, the sample changer cannot read the RFID
tag of the turntable. Check that a turntable is correctly mounted on the
sample changer.

Check the model of turntable used and refer to the User’s Guide of the
sample changer, chapter 6 "Accessories".

Solve the problem then restart the sequence.

TTL 5 V OUT
(sockets)

Red and black banana sockets (diameter = 2 mm).

The red banana sockets send a TTL 0 ± 5 V.

The black banana sockets are connected to the electrical zero of the
instrument.

This auxiliary signal is programmed in the Method parameter or Rea-
gent parameter menus,

see "Auxiliary output", page 42

.

Specifications (potential difference is given in absolute value)

Output signal

Level 0: 0 to 0.4 V

Level 1: higher than 2.4 V and equal to or less than 5 V

Output impedance: 1 KOhm at level 0 and 2 kOhm at level 1

A TTL 0-level does not mean a 0 V output.
The voltage difference varies from 0 to ±400 mV.