Burkert Type 8045 User Manual
Page 62

58
Operatingandfunctions
FLOw-w.
END
w-= 0.000
w+= 0.000
→
Enter	a	flow	rate	value	(in	the	unit	chosen	in	the	"UNIT"	parameter)	
such	as	W+	>	W-,	above	which	a	"WARN	HI"	warning	message	is	
generated	by	the	device.
→
Confirm.
→
Enter	a	flow	rate	value	(in	the	unit	chosen	in	the	"UNIT"	parameter),	
below	which	a	"WARN	LO"	warning	message	is	generated	by	the	
device.
→
Confirm.
Fig. 60:
Diagram of the "FLOW-W." parameter of the Test menu
To	be	warned	when	the	flow	rate	is	too	low	or	too	high,	parameter	the	flow	rate	range	(in	the	units	that	have	
been	chosen	in	the	"UNIT"	parameter	of	the	Parameters	menu),	outside	which	the	device	generates	a	warning	
message,	"WARN	LO"	or	"WARN	HI",	and	turns	the	device	status	LED	to	orange.
When a warning message, "WARN LO" or "WARN HI", is generated by the device:
→
check the process.
→
if the process is not faulty, check the flow sensor condition and clean it if necessary.
→
if the flow rate measurement is still faulty, contact the Bürkert retailer.
• The transistor output or either relay output can be configured to switch when a warning message is gen-
erated by the device. See chap. 8.6.5.
• See also "If you encounter problems" in chap. “9.5”
→
If	you	do	not	want	to	adjust	another	parameter,	confirm	the	"END"	parameter	to	save	the	settings	or	not	and	
go	back	to	the	Process	level.
English
Type 8045
