beautypg.com

Figure 5-14, Figure 5-16, Figure 5-15 – Atec Tektronix-1502 User Manual

Page 68: Figure 5-17, Install fuseholder caps (see fig. 5-14)

background image

TM 9-4935-601-14-3&P

Figure 5-14. Detail of Fuseholder Caps.

Figure 5-16. Alignment of Battery Meter

3. Insert the caps is the Fuseholders and press to ensure the washers are fully seated.
4. Remove the caps, inspect, and remove any excess sealant.
5. Apply a film of G-661 lubricant to exposed surface of washer.
6. Insert fuse (see Electrical Parts list for correct value) into cap and install cap with fuse in the

Fuseholder.

16. Install Battery Meter (see Figs. 5-15 and 5-16)

a. METER MPL Item 2-53.

Figure 5-15. Application of sealant to Battery Meter.

Figure 5-17. Application of sealant to FOCUS
and INTENSITY controls.

b. Procedure

1. Apply a thin bead of No. 3144 sealant to the junction of the bezel and meter canister (see Fig. 5-15).
2. Install Meter in Front Panel with markings aligned as shown in Fig. 5-16.

17. Install the FOCUS and INTENSITY controls (see Fig. 5-17)

a. RESISTOR, VARIABLE (2 ea.). MPL Item 2-38 and Item 2-41

These two variable resistors are used for FOCUS and INTENSITY control and are listed in the Electrical Parts
list.

b. Procedure

1. Unscrew the bushing part of the resistor assembly and pack the barrel/shaft interface with G-661

lubricant.

@FEB 1980

5-12