Atec Tektronix-1502 User Manual
Page 13
TM 9-4935-601-14-3&P
Table 1-2. (cont)
ELECTRICAL CHARACTERISTICS
Characteristics
Performance Requirements
Supplemental Information
Line Voltage
117 Vac ±20%, 48 to 410 Hz
Fused at 0.5 A
234 Vac
±
20%, 48 to 410 Hz
Fused at 0.3 A
Battery Pack
C size 9 cell
Operation
At least 5 hours
+20° C to +25° C
charge and discharge temperature
Full Charge Time
16 hours
Typical Charge Capacity
Charge Temperature
Discharge Temperature
-15°C
+20°C to
+55°C
+25°C
0° C
40%
60%
50%
+20° C to +25° C
65%
100%
85%
+40° C
40%
65%
55%
Temperature
Operating
-15° C to +55° C
At temperatures other than 20°C-
25° C, the battery efficiency
becomes restricted
Non-Operating
-62° C to +85° C
With batteries removed. If stored
with batteries storage range
changed to -40° C to +55° C
Humidity
To 100%
Altitude
Operating
10,000 feet
Non-Operating
50,000 feet
Vibration
3.0 g, 5 to 55 Hz; Test time
45 minutes
Shock, Mechanical
Shock, Pulse
15 g, 1/2 sine shock waveform of
11 ms duration. Total of 18 shocks
Bench Handling
Operating
4 drops each face at 4 inches or 45°
Case on
with opposite edge as pivot
Non-Operating
4 drops each face at 4 inches or 45°
Case off
with opposite edge as pivot.
Satisfactory operation after drops.
Transit Drop
12 inch drop
REV B FEB 1980
1-4