TM 9-4935-601-14-3&P

Adjustment Procedure

58. Set the FEET/DIV multiplier to X1.
59. Set the incident pulse to a graticule reference line with the ZERO REF control, then set the FEET/DIV

multiplier to X.1.

60. Check-that the leading edge of the incident pulse is located within 1 division of the selected graticule

reference line.

61. Adjust-R1132 (X.1 POS CAL) located on the MAIN board, so that the leading edge of the incident pulse

is located on the selected graticule reference line. Repeat steps 58 through 60.

HORIZONTAL CHECKS AND ADJUSTMENTS-OPTION 5, METRIC VERSION

(See Step 38 for Standard Version)

Horizontal Gain Presets

NOTE

If the unit is only being checked and not calibrated, or if the unit is to be used
with the optional chart recorder, do not perform the following Horizontal Gain
Presets procedure. Instead, turn to the metric version Timing procedure
beginning with step 67.

62. Set the mp/DIV to 500.
63. Use the POSITION controls to center the trace on the graticule.
64. Set R3217 (HORIZ) located on the OUTPUT AMPS board, to the counterclockwise stop.
65. Adjust-R3213 (H POS) located on the OUTPUT AMPS board, to center the trace on the graticule.
66. Adjust-R3217 (HORIZ) so that the retrace at the start of the sweep is just off the graticule area and the

trace covers all 10 divisions of the graticule.

NOTE

Once the horizontal gain has been set, do not change it.

Timing

67. Connect the Marker Out of the TG 501 through a coaxial cable and a 10X attenuator to the CABLE

connector of the 1502; also connect the TG 501 Trigger Out to the Trig/Duration Input of the PG 502.

Turn the instrument POWER off. Turn the instrument upside down and remove the shield that covers the

pulser/sampler circuitry (remove 12 screws). Remove U1575 from its socket. In the following steps,
instructions are given on how to provide an external trigger to the 1502 (see Fig. 4-6).

68. Set the 1502 controls as follows:

mp/DIV

200

METRES/DIV

25 @ X1

DIELECTRIC

AIR (all buttons out)

DISTANCE

000

69. Set the controls of the PG 502 Pulse Generator as follows:

Duration

50

µ

s

Variable Duration

Fully ccw

Period

Ext Trig

Low Level

-1.5

High Level

2.5

Back Term

Pull Out

All pushbuttons

Out

70. Set the TG 501 Time Mark Generator for 0.5 us time marks. Connect a coaxial cable from the TG 501

+ Trigger Out to PG 502 Trig/Duration Input. Connect a 1X probe from the PG 502 Output to the Red test point
located near U1575. Connect a coaxial cable and a 10X attenuator from the TG 501 Marker Out to the CABLE
connector. Turn the instrument POWER on.

NOTE

The time marks must not exceed 0.85 V to keep the tunnel diode in the low
state.

71. (a). Obtain the waveform on the 1502 crt as shown in Fig. 4-9. Use the variable pulse amplitude

control of the PG 502 and the DISTANCE control of the 1502 to align the time marks with the crt graticule lines.

NOTE

If the instrument is to be used with the optional chart recorder, continue with the
remainder of this step. If the instrument is not to be used with a chart recorder,
proceed with step 72, part a.

REV C FEB 1980

4-12