Figure 2-3 – Atec Tektronix-1502 User Manual
Page 18
TM 9-4935-601-14-3&P
Figure 2-3. 1502 Front Panel Controls.
7.
GAIN
Screwdriver adjust to set the gain
13.
DISTANCE
Indicates the distance from the
of the vertical amplifier.
1502 to the point on the cable
where the display window begins.
Two ranges: 100 feet (25 m) at X.1
8.
NOISE FILTER
Reduces displayed noise. Display
or 1000 feet (250 m) at X1.
rate is reduced by a factor of 10.
Disabled when the FEET/DIV
(METRES/DIV) is at 200 (FIND)
(50).
9.
BATTERY
Meter to indicate the relative
charge of the power pack.
14.
FEET/DIV
Selects the horizontal deflection
10.
ZERO REF
Momentary contact pushbutton.
(METRES/DIV)
factor:
CHECK
When pushed, checks the horizon-
X1 = 1 - 200 ft/div (25 cm -
tal location of the incident pulse on
50 m/div).
the crt when the DISTANCE dial is
being used.
X.1 = 0.1 - 20 ft/div (2.5 cm -
5 m/div).
11.
ZERO REF SET
Horizontal pulse position control
for crt display. Sets the incident
pulse edge to a vertical reference
line of the crt when the DISTANCE
15.
CABLE
Three pushbuttons and a screw-
dial is at 000 or the ZERO REF
DIELECTRIC
driver adjust. Selects the proper
CHECK button is at pushed.
SOLID POLY
velocity of propagation. VAR from
0.55 to 1.0 when the OTHER
SOLID PTFE
pushbutton is pressed. Fully CW is
12.
MULTIPLIER
Two-position switch (red control)
for air dielectric. VAR control has
for X.1 or X1 multiplier. Affects
OTHER VAR
reference marks every 30° to in-
both the DISTANCE dial and the
dicate relative propagation con-
FEET/DIV (M ETRES/DIV) control.
stants.
REV B FEB 1980
2-3