Section iv – Atec Tektronix-1502 User Manual
Page 39
TM 9-4935-601-14-3&P
CALIBRATION PROCEDURE
The first part of this procedure checks the electrical characteristics of the 1502 that appear in the
Specification section of this manual. This procedure can also be used by an incoming inspection facility to
determine acceptability of performance. The second part of this section contains a detailed adjustment
procedure.
It is recommended that these elementary checks be performed before those in the DETAILED CHECKS
AND ADJUSTMENTS section; these checks will identify most conditions calling for troubleshooting and repair.
To aid in troubleshooting, the probable problem area(s) are given after each applicable step. Refer to the Block
Diagram and Schematics in Section 8 as an aid in locating components, test points and circuits.
Table 4-1.
TEST EQUIPMENT REQUIRED
Performance
Description
Requirement
Application
Example
Precision 50
Ω
Cable (2)
Known length and
Timing
Tektronix part
impedance
number 012-0482-00
Precision 50
Ω
Known impedance
Gain and impedance
Tektronix part
Terminator
checks
number 011-0123-00
BNC female-to-female
Calibration
Tektronix part
adapter
number 103-0028-00
BNC male-to-probe
Calibration
Tektronix part
adapter
number 013-0130-00
BNC-to-GR adapter
Known impedance
Calibration/Performance
Tektronix part
Check
number 017-0064-00
GR short
Low inductance
Calibration/Performance
Tektronix part
Check
number 017-0087-00
Power Module
Calibration
TEKTRONIX TM 503
Power Module
Digital Multimeter
0-200 V dc
Calibration
TEKTRONIX DM 501
Time Mark Generator
5
µ
s to 10
µ
s markers
Calibration
TEKTRONIX TG 501
Pulse Generator
50
µ
s period
Calibration
TEKTRONIX PG 502
2.5
µ
s pulse
Var dc level
5 V pulse
1 X Probe
Calibration
TEKTRONIX P6101
REV B FEB 1980
4-1