7
E8362/3/4B
Corrected system performance with 2.4 mm connectors
Description
Specification (dB)
45 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
Directivity
8
6
Source match
1
8
1
Load match
7
5
Reflection tracking
±0.001 (+0.0/°C)
±0.008 (+0.0/°C)
±0.00 (+0.0/°C)
±0.07 (+0.0/°C)
Transmission tracking
±0.010 (+0.0/°C)
±0.09 (+0.0/°C)
±0.105 (+0.0/°C)
±0.170 (+0.0/°C)
Transmission uncertainty (specifications)
Reflection uncertainty (specifications)
Magnitude
0.01
0.1
1
10
-90
-80
-70
-60
-50
-0
-0
-0
-10
0
10
Transmission coefficient (dB)
Uncertainty (dB
)
E86/B full two port cal using 85056A
S11 = S = 0
Source power = -1 dBm
5 MHz to GHz
to 0 GHz
0 to 0 GHz
0 to 50 GHz
Phase
0.1
1
10
100
-90
-80
-70
-60
-50
-0
-0
-0
-10
0
10
Transmission coefficient (dB)
Uncertainty (degrees)
E86/B full two port cal using 85056A
S11 = S = 0
Source power = -1 dBm
5 MHz to GHz
to 0 GHz
0 to 0 GHz
0 to 50 GHz
Magnitude
0
0.01
0.0
0.0
0.0
0.05
0
0.
0.
0.6
0.8
1
Reflection coefficient (linear)
Uncertainty (linear)
E86/B with 85056A
S1 = S1 = 0
Source power = -1 dBm
5 MHz to GHz
to 0 GHz
0 to 0 GHz
0 to 50 GHz
Phase
0
6
8
10
0
0.
0.
0.6
0.8
1
Reflection coefficient (linear)
Uncertainty (deg)
E86/B with 85056A
S1 = S1 = 0
Source power = -1 dBm
5 MHz to GHz
to 0 GHz
0 to 0 GHz
0 to 50 GHz
Standard configuration and standard power range (E8363/4B)
Applies to E8363/4B PNA Series analyzer,
85056A (2.4 mm) calibration kit, 85133F flexible
test port cable set, and a full two-port calibration.
(Specifications apply over environmental temperature
of 23 °C ±3 °C, with less than 1 °C deviation from