beautypg.com

E8361a, Test port input – Atec Agilent-E8361A User Manual

Page 39

background image

9

E8361A

Test port input

continued

Description

Specification

Option 014 or

Supplemental

Standard

Option 014 and UNL (typ.)

information

Direct receiver access

input noise floor

1

10 Hz IF bandwidth

10 to 5 MHz

< -106 dBm

5 to 500 MHz

, 5

< -105 dBm

Online Help also includes the

500 MHz to GHz

< -15.5 dBm

category “Direct receiver access

to 10 GHz

< -15 dBm

input noise floor, Option 080 enabled”

10 to GHz

< -18 dBm

to 0 GHz

< -117.5 dBm

0 to 0 GHz

< -117 dBm

Option 016 degrades

0 to 5 GHz

< -115 dBm

performance by dB.

5 to 50 GHz

< -11.5 dBm

50 to 60 GHz

< -111 dBm

60 to 67 GHz

< -108 dBm

Option 016 degrades

67 to 70 GHz

< -107 dBm

performance by dB.

1 kHz IF bandwidth

10 to 5 MHz

< -86 dBm

5 to 500 MHz

, 5

< -85 dBm

500 MHz to GHz

< -105.5 dBm

to 10 GHz

< -105 dBm

10 to GHz

< -108 dBm

to 0 GHz

< -97.5 dBm

0 to 0 GHz

< -97 dBm

Option 016 degrades

0 to 5 GHz

< -95 dBm

performance by dB.

5 to 50 GHz

< -9.5 dBm

50 to 60 GHz

< -91 dBm

60 to 67 GHz

< -88 dBm

Option 016 degrades

67 to 70 GHz

< -87 dBm

performance by dB.

Description

Specification

Supplemental

Receiver compression level (measured at Test Ports)

Supplemental

Specifications

Information

Standard

Option 014

Option 014 and UNL

Typical

10 to 5 MHz

,

negligible

negligible

negligible

negligible

5 to 500 MHz

, 6

<0.1 dB at -9.5 dBm

7

<0.1 dB at -9.5 dBm

7

<0.1 dB at -9.5 dBm

7

<0.1 dB at +0.5 dBm

7

and <0.5 dB at - dBm

and <0.5 dB at - dBm

and <0.5 dB at - dBm

and <0.5 dB at +8 dBm

500 MHz to 5 GHz <0.1 dB at -8 dBm

7

<0.1 dB at -8 dBm

7

<0.1 dB at -7 dBm

7

<0.1 dB at - dBm

7

and <0.5 dB at -1 dBm

and <0.5 dB at -1 dBm

and <0.5 dB at 0 dBm

and <0.5 dB at + dBm

5 to 0 GHz

<0.1 dB at -8.5 dBm

7

<0.1 dB at -8.5 dBm

7

<0.1 dB at -6 dBm

7

<0.1 dB at -1 dBm

7

and <0.5 dB at - dBm

and <0.5 dB at - dBm

and <0.5 dB at +1 dBm and <0.5 dB at +6 dBm

0 to 67 GHz

< 0.1 dB at -10.5 dBm

7

< 0.1 dB at -8 dBm

7

<0.1 dB at –9.5 dBm

7

<0.1 dB at - dBm

7, 8

and <0.15 dB at -7 dBm

and <0.15 dB at - dBm

and <0.15 dB at -6 dBm

and <0.15 dB at + dBm

8

67 to 70 GHz

<0.1 dB at - dBm

7, 8

and <0.15 dB at + dBm

8

1. Total average (rms) noise power calculated as mean value of a linear magnitude

trace expressed in dBm.

. Typical performance.

. Coupler roll-off will reduce compression to a negligible level below 500 MHz.

. Noise floor may be degraded by 10 dB at particular frequencies

(multiples of 5 MHz) due to spurious receiver residuals.

5. Specified value is for worst-case noise floor at 5 MHz.

6. Specified value is for worst-case compression at 500 MHz.

7. This compression level comes from the dynamic accuracy curve with -0 dBm

reference test port power.

8. Option 016 degrades performance by dB.