beautypg.com

Corrected system performance, System dynamic range – Atec Agilent-E8361A User Manual

Page 5

background image

5

E8362/3/4B

Corrected system performance

continued

1. The system dynamic range is calculated as the difference between the noise

floor and the source maximum output power. System dynamic range is a

specification when the source is set to port 1, and a characteristic when the

source is set to port . The effective dynamic range must take measurement

uncertainties and interfering signals into account.

. The test port system dynamic range is calculated as the difference between the

test port noise floor and the source maximum output power. The effective

dynamic range must take measurement uncertainties and interfering signals

into account.

. The direct receiver access input system dynamic range is calculated as the

difference between the direct receiver access input noise floor and the source

maximum output power. The effective dynamic range must take measurement

uncertainties and interfering signals into account. This set-up should only be

used when the receiver input will never exceed its damage level. When the

analyzer is in segment sweep mode, the analyzer can have pre-defined

frequency segments which will output a higher power level when the extended

dynamic range is required (i.e. devices with high insertion loss), and reduced

power when receiver damage may occur (i.e. devices with low insertion loss).

The extended range is only available in one-path transmission measurements.

. Typical performance.

5. May be limited to 100 dB at particular frequencies below 500 MHz due to

spurious receiver residuals. Methods are available to regain the full dynamic

range.

6. E86B only: Option H11 decreases value by 1 dB.

7. E86B only: Option H11 decreases value by dB.

System dynamic range

1

Description

Specification (dB) Typical (dB) at direct

Supplemental information

at test port

2

receiver access input

3

Dynamic range
Standard configuration and extended power range and bias-tees (E8362/3/4B-Option UNL)
10 to 5 MHz

79

N/A

5 to 500 MHz

5

9

N/A

500 MHz to GHz

117

N/A

to 10 GHz

10

N/A

10 to 0 GHz

11

N/A

0 to 0 GHz

11

N/A

0 to 0 GHz

108

N/A

Option 016 degrades

0 to 5 GHz

105

N/A

performance by dB

5 to 50 GHz

99

N/A

Configurable test set and extended power range and bias-tees

(E8362/3/4B-Option UNL and Option 014)
10 to 5 MHz

79

19

5 to 500 MHz

5, 6

9

10

500 MHz to GHz

6

117

16

to 10 GHz

6

10

15

10 to 0 GHz

7

119

1

0 to 0 GHz

109

11

0 to 0 GHz

105

117

Option 016 degrades

0 to 5 GHz

101

11

performance by dB

5 to 50 GHz

95

106