E8361a, Test port output – Atec Agilent-E8361A User Manual
Page 37
![background image](/manuals/318349/37/background.png)
7
E8361A
Test port output
continued
Description
Specification
Supplemental information
Phase noise (10 kHz offset from center frequency, nominal power at test port)
10 to 5 MHz
80 dBc typical
5 MHz to 10 GHz
70 dBc typical
10 to GHz
60 dBc typical
to 70 GHz
55 dBc typical
Phase noise (10 kHz from center frequency, nominal power at test port) – Option 080 enabled
10 to 5 MHz
80 dBc, typical
5 MHz to 10 GHz
70 dBc, typical
10 to GHz
60 dBc, typical
to 70 GHz
55 dBc, typical
Phase noise (100 kHz from center frequency, nominal power at test port)
10 to 5 MHz
90 dBc, typical
5 MHz to 10 GHz
90 dBc, typical
10 to GHz
85 dBc, typical
to 70 GHz
75 dBc, typical
Phase noise (100 kHz from center frequency, nominal power at test port) – Option 080 enabled
10 to 5 MHz
85 dBc, typical
5 MHz to 10 GHz
80 dBc, typical
10 to GHz
70 dBc, typical
to 70 GHz
60 dBc, typical
Phase noise (1 MHz from center frequency, nominal power at test port)
10 to 5 MHz
115 dBc, typical
5 MHz to 10 GHz
110 dBc, typical
10 to GHz
105 dBc, typical
to 70 GHz
95 dBc, typical
Phase noise (1 MHz from center frequency, nominal power at test port) – Option 080 enabled
10 to 5 MHz
110 dBc, typical
5 MHz to 10 GHz
105 dBc, typical
10 to GHz
95 dBc, typical
to 70 GHz
85 dBc, typical
Harmonics (nd or rd)
10 to 500 MHz
10 dBc typical, in power
500 MHz to 10 GHz
15 dBc typical, in power
10 to GHz
dBc typical, in power
to 50 GHz
16 dBc typical, in power
50 to 60 GHz
1 dBc typical, in power
60 to 70 GHz
19 dBc typical, in power
Non-harmonic spurious (at nominal output power)
10 MHz to 0 GHz
-50 dBc typical, for offset
0 MHz to 70 GHz
frequency > 1 kHz
-0 dBc typical, for offset
frequency > 1 kHz