beautypg.com

Stp siphon flow rate test – Franklin Fueling Systems TS-550/TS-5000 consoles Secondary Containment Monitoring User Manual

Page 19

background image

16

STP Siphon Flow Rate Test

Warning

Failure to open the leak generator (TS-SCMCAL) during this test can apply an excessive vacuum

level to the containment, causing irreversible damage. Ensure that the leak generator is installed

properly and that the toggle switch is in the OPEN position prior to energizing the STP.

Note:  During this procedure the containment reaches 9" Hg, which will cause the console to display a High Vacuum 

alarm and disable the containment. If Pump Shutdown on Alarm is chosen in the options, a disabled containment 

will not allow the STP to be activated. Navigate to the SCM Control screen, press 

Reset Alarm to continue.

1.  Install a vacuum gauge and the vacuum leak generating device (TS-SCMCAL) into the leak generator (LG) port on the 

SCCM using necessary fittings. Switch the toggle valve to the UP (open) position.

2.  At the T5 series FMS console (using either the LCD or Tank Sentinel Anyware), navigate to the SCM Control Menu. 

Press the 

Solenoid button — this activates (opens) the solenoid. 

3.  Activate the STP by lifting the appropriate dispenser handle. 
4.  With the STP on and the toggle valve open, the vacuum level must be able to reach at least 6.5" Hg. Verify vacuum 

levels at the gauge connected to the containment area. 
Large Containments (>50 gal.) may take several minutes. For large containments, close the leak generator until the 

vacuum level reaches approximately 6" Hg. Open the leak generator. Allow the vacuum to build. The vacuum level 

must be able to reach at least 6.5" Hg, but, if the vacuum is not able to reach at least 6.5" Hg, then the STP siphon 

assembly must be repaired or replaced.

5.  Once the vacuum level is verified above 6.5" Hg, return the hook signal to turn off the STP. 
6.  At the T5 series FMS console (LCD or TSA), navigate to the SCM Control Menu. Press the 

Solenoid button — this 

deactivates (closes) the solenoid.

7.  Allow the vacuum to decay to 0" Hg.

If you are proceeding to the Learn Process chapter next, the TS-SCMCAL may remain installed at the SCCM leak 

generator port.