Xylem SuSix Turbidity and Suspended Solids Transmitter User Manual
Page 96

96
EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Sensor
Error - General sensor error that
may also be transferred to a relay
output.
Contact MJK.
Sensor communication
Error - Check the cable condition
and tighten the connectors.
Sensor hardware
Error - Change the sensor hard-
ware.
Sensor setting
Error - Some settings are wrong or
have been changed.
Sensor supply
Error - The sensor supply is too
low or too high. Check the sensor
cable connection.
Sensor wiper
Error - Make sure that the wiper is
free to move and not obstructed by
dirt deposits.
System
Error - Internal VCC error. See the
event log for details.
System error
Error - The system software is in-
operable including faulty alarm and
sensor bits.
Contact MJK.
Turbidity offset
Error - A zero-point probe prob-
lem. The sensor needs a zero
calibration.
Contact MJK.
Warnings
Needs higher meas.sample
Warning - Calibrate at a higher
concentration.
Needs lower meas.sample
Warning - Calibrate at a lower
concentration.
Sensor range exceeded
Warning - The measured value is
out of range. Check the measuring
range and the measuring units.