Turbidity and Suspended Solids Transmitter

Sensor

Error - General sensor error that
may also be transferred to a relay
output.
Contact MJK.

Sensor communication

Error - Check the cable condition
and tighten the connectors.

Sensor hardware

Error - Change the sensor hard-
ware.

Sensor setting

Error - Some settings are wrong or
have been changed.

Sensor supply

Error - The sensor supply is too
low or too high. Check the sensor
cable connection.

Sensor wiper

Error - Make sure that the wiper is
free to move and not obstructed by
dirt deposits.

System

Error - Internal VCC error. See the
event log for details.

System error

Error - The system software is in-
operable including faulty alarm and
sensor bits.
Contact MJK.

Turbidity offset

Error - A zero-point probe prob-
lem. The sensor needs a zero
calibration.
Contact MJK.

Warnings

Needs higher meas.sample

Warning - Calibrate at a higher
concentration.

Needs lower meas.sample

Warning - Calibrate at a lower
concentration.

Sensor range exceeded

Warning - The measured value is
out of range. Check the measuring
range and the measuring units.