Turbidity and Suspended Solids Transmitter

EN 5.10 SuSix Manual 1406

Analogue Output

The analogue output is an active output with a maximum load of 800Ω.

Analogue Output

Terminal

Designation

AO

+

4-20 mA

AO

-

4-20 mA

The analogue output can be programmed for indication of primary units.
See details on page

61

.

Digital Output

SuSix

®

has two digital outputs - DO1 with an opto (light triggered) relay

and DO2 with an electro-mechanical relay.
They can both be programmed for the following functions:

-

Sensor error

-

High and low alarms

Opto Relay

Terminal

Designation

DO1

Com

Max. 50 V DC /120 mA

DO1

NO

Electro-mechanical relay

Terminal

Designation

DO2

Com

Max. 50 V DC /1 A

DO2

NO

DO1 shares the common terminal (Com) with DI.
See also application examples in Appendix E. Digital Input/Output Con-
nections on page

111

.