Appendix a. pop-up and error messages – Xylem SuSix Turbidity and Suspended Solids Transmitter User Manual
Page 95

95
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
EN 5.10 SuSix Manual 1406
Appendix A. Pop-up and Error Messages
Alarms
Current overrange
Alarm - The mA connection has
exceeded the 4 - 20 mA range. For
example 3,5 or 20,5 mA.
Factory reset occurred
Alarm - A factory reset has oc-
cured. Some settings have
changed. Check all used settings.
High alarm
Alarm - The high set point has
been detected.
Low alarm
Alarm - The low set point has been
detected.
mA connection
Alarm - The mA connection is not
connected. If the current-out is
used, deactivate mA Current out to
stop this alarm.
Errors
Calibration
Error - 1. Perform calibration (see
page
99
) and/or check the units
and ranges (see page
60
).
Eeprom write error
Error - Eeprom write error. If error
keeps popping up, contact MJK.
Measuring start failed
Error - Start failed. Try again.
Measuring timeout
Error - Timeout. Try again.
Missing adjustment
Error - The measured value has
not been entered.
Not in solid mode!
Error - Calibration can only be
performed when in solid mode.
Change mode and try again.