beautypg.com

Required tools and materials – Xylem SuSix Turbidity and Suspended Solids Transmitter User Manual

Page 20

background image

20

EN 5.10 SuSix Manual 1406

FW: 841013-005/843003-001

Turbidity and Suspended Solids Transmitter

Installation / Removal using Insertion Fitting 155210

Warnings: Max. operating pressure is 6 bar (720 N).

Max. residual pressure when changing sensor is 1 bar (120 N).

The sensor guide (2) is not a safety component

The insertion fitting is not fully assembled from the factory

Be careful when dealing with corroding and toxic substances

Be careful when dealing with lines under pressure

Required Tools and Materials

• 5 mm hex key
• LOCTITE

®

577 sealant

T

7

1

2

3

4

5

6

Measure position Service position

A

B

Preparing the Insertion Fitting before welding to tube

The nuts on the ball valve housing (see A and B arrows) have not been
tightened and secured with a sealant from the factory. This allows the
customer to rotate, position and secure the ball valve as required on the
installation site.
1. Loosen/remove the A and B nuts, and rotate the ball valve to its final

position.