beautypg.com

Maintenance, Cleaning the sensor - once a month, Cleaning the transmitter – Xylem SuSix Turbidity and Suspended Solids Transmitter User Manual

Page 89: Functional tests - once a year

background image

89

FW: 841013-005/843003-001

Turbidity and Suspended Solids Transmitter

EN 5.10 SuSix Manual 1406

10. Maintenance

The SuSix

®

sensor should be cleaned on a regular basis to maintain its

accuracy.
Initially the the transmitter should be cleaned once a week or every two
weeks until the optimal cleaning interval can be determined.

Cleaning the Sensor - Once a Month

1. Remove the transmitter from the liquid.
2. Clean the sensor with a soft, moist cloth and clean water. A mild deter-

gent or dishwashing liquid may be used.

3. Dry the sensor with a soft cloth.
4. Immerse the transmitter into the liquid to be measured.
Note: The accuracy will deteriorate, if the measuring optics are

scratched by for example a steel brush during cleaning.

5. The cleaning intervals may need to be shortened all depending on the

usage and the location.

Cleaning the Transmitter

When seen necessary the SuSix

®

transmitter may be cleaned with a soft,

moist cloth.

Functional Tests - Once a Year

Electronics

1. Check the suspended solids transmitter calibration.
2. Tighten the terminal screws and ensure that all leads are connected

correctly.

3. A service contract maintains the equipment reliability and accuracy. Get

in contact with MJK’s national sales and service representatives.