Xylem SuSix Turbidity and Suspended Solids Transmitter User Manual

Page 82

background image

82

EN 5.10 SuSix Manual 1406

FW: 841013-005/843003-001

Turbidity and Suspended Solids Transmitter

Change device address
To change each converter's device adress it is required that you connect
the Display Unit to each converter one by one to change each converter's
device address. If neglected, a conflict between the units will arise when
the RS 485 serial loop is established.
Change for example "Device 1" to "Device 2" as follows:
1. Press OK and use the up/down keys to change "1" to "2".
2. Press OK.
The SuSix

®

converter ID is now changed to "2", and more converters can

be added to the Display Unit with the "Add Device" option.

Remove Device
If a device is to be removed from the Display Unit, proceed as follows:
1. Select the required device with the up/down keys.
2. Click OK to remove it.
The device is now removed from the Display unit and display returns to the
"Modbus" menu.

Change controller ID (display)
The display controller ID must be changed when several displays operate
with the same converter(s)