5 reactivating the last valid calibration – YSI P 700 IQ Analyzer User Manual
Page 62
62
ba76133e01
09/2013
P 700 IQ
• t (offset)
• Calibration evaluation:
• +: Successful calibration.
The new calibration data are taken over for measurement.
• -: Calibration unsuccessful.
Sensor blocked for measurement.
4.2.5 Reactivating the last valid calibration
If a manual calibration procedure is unsuccessful, the measuring oper-
ation can only be resumed after the manual calibration was repeated or
a valid calibration was reactivated.
If an automatic calibration procedure is unsuccessful, the analyzer
stops. To be able to measure again, you have to carry out a valid cali-
bration procedure in the calibration and service menu, or to reactivate
a valid calibration.
Thus you can immediately continue to measure if a calibration failed or
you suspect that the calibration conditions were not optimally met.
1
Using <>, select the P 700 IQ in the measured value display.
2
Using
The maintenance condition is activated.
3
Confirm the maintenance condition with
The calibration and service menu is displayed.
4
Carry out the Start measuring function.
A menu to select the last valid calibration or the factory calibration
is displayed.
5
Select the last valid calibration or the factory calibration.
6
Carry out the Exit SERVICE function.
or
Exit the calibration and service menu with
7
Confirm the note with
appears. The measured value flashes.
8
If necessary, disconnect the terminal from the MIQ/WCA 232.
9
Close the enclosure of the P 700 IQ.
Reactivating old calibration data is a temporary measure.
Take into consideration that the sensor my provide wrong
measured values. Ensure the correct functioning of the
sensor by checking and/or recalibrating it.