Insert the slide of the filter membrane module in the rail and lower
it into the basin with the aid of the chain. Fix the end of the chain
outside the basin.
5
Run the suction line to the analyzer. Fix the line with cable ties at
some suitable places as necessary. How to connect the suction
line to the P 700 IQ, see section 3.3.4, 36.
3.3.4 Inserting the connections into the housing
Connection plate
All connections are inserted into the housing through dust-proof lead-in
ducts. The lead-in ducts are on a connection plate in the bottom of the
housing. The connection plate can be removed for the mounting of the
tubes.
1
Unscrew the connection plate from the bottom of the housing.
2
Run the power line through the small cable gland on the left. Fix
the power line only slightly for the time being so that it can be
readjusted later.
3
Run the IQ S
ENSOR
N
ET
cable through the small rear cable gland.
Fix the IQ S
ENSOR
N
ET
cable only slightly for the time being so
that it can be readjusted later.
Connecting the filter
unit and return tube
(accessories)
4
Run the suction line through the big cable gland on the left side
of the connection plate.
• The sleeve tube should stick out approx. 5 cm behind the con-
nection plate.
• Fix the sleeve tube with the cable gland.
5
Run the return tube through the big cable gland on the right side
of the connection plate.
• The sleeve tube should stick out approx. 5 cm behind the con-
nection plate.
• Fix the sleeve tube with the cable gland.
If the IQ S
ENSOR
N
ET
should be further branched at the
MIQ/WCA 232, insert the IQ S
ENSOR
N
ET
cable leading
away through the small front cable gland.