4.2 Calibration
4.2.1 Overview
Why calibrate?
During operation, components of the P 700 IQ can age or become dirty
and thus change their characteristics, e.g.:
• Flow-through cell
• LED
• Photo diodes
• Color reagent
Regular automatic or manual calibration procedures help you to recog-
nize any changes of the measuring characteristics of the P 700 IQ.
When to calibrate?
Calibrate at regular intervals.
1-point calibration
The offset of the characteristic curve can be adjusted with a regular 1-
point calibration.
The 1-point calibration can be carried out automatically at regular inter-
vals. An additional 1-point calibration procedure can be started in the
calibration and service menu at any time. See section 4.2.2, 55.
The Autom.cleaning, Autom.calibration and Meas. interval
settings may result in the overlapping of the carrying out of
different functions.
In this case, the functions are carried out according to pri-
ority.
Function
Priority
Duration
(min)
Measuring
range
Autom.cleaning
1
5
A, B
Autom.calibration
2
6
9
A
B
Measurement
3
5
7
A
B