beautypg.com

YSI P 700 IQ Analyzer User Manual

Page 45

background image

ba76133e01

09/2013

45

P 700 IQ

6

As soon as enough sample is running, reduce the pump power to
5 - 10 % with the aid of the arrow keys.

7

Close the doors of the enclosure.

8

If necessary, commission the IQ S

ENSOR

N

ET

system (see the rel-

evant IQ S

ENSOR

N

ET

system operating manual).

If no sample runs into the overflow vessel, proceed in one
of the following ways:

• Fill the suction line with water manually:

• Switch off the filtration pump.

• Pull the intake tube off the filtration pump.

• Fill the intake tube with water (e.g. using a wash

bottle).

• Connect the intake tube to the pump again and

secure it.

• Switch on the filtration pump.

• Increase the suction power manually:

• At the vent connection (2), connect a syringe with

a tube:

• Open the vent valve (1).

• Create an additional negative pressure with the

syringe.

• Close the vent valve (1).

• Repeat the procedure until sample runs into the

overflow vessel.

• Remove the tube and syringe from the vent con-

nection (2).

Fig. 3-21: Filtration pump

1 Vent valve
2 Vent connection

1

2