2 info messages, 2 status info, Status info – YSI P 700 IQ Analyzer User Manual
Page 107

ba76133e01
09/2013
107
P 700 IQ
8.1.2 Info messages
The analyzer does not generate any info messages.
8.2 Status info
The status info is a coded piece of information on the current status of
a sensor. Each sensor sends this status info to the controller. The
status info of sensors consists of 32 bits, each of which can have the
value 0 or 1.
IC53Bx
P700IQ-PO4 was successfully calibrated
* For calibration data, see calibration history
II13Bx
Language not available,
Default language German
* Contact service
IC63Bx
Liquid tank xxx almost depleted!
* Please refill to maintain proper functioning
EI155x
Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components, replace components
EI255x
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components,
replace components
II155x
Language not available,
Default language German
* Contact service
ES255x
Component hardware defective xxx
* Contact service
EIA55x
Communication between MIQ/WCA and P700IQ-PO4 disturbed
* Check cable connections
* Provide voltage supply to P700IQ-PO4
* Contact service
Message code