3 installing the fm filter membrane module and the, M 1.5 attachment for filtration, See section 3.3.3, 35 – YSI P 700 IQ Analyzer User Manual
Page 35
ba76133e01
09/2013
35
P 700 IQ
3.3.3 Installing the FM filter membrane module and the
M 1.5 attachment for filtration
Installation instructions
Heed the following notes when installing the filter membrane module:
• Mount the filter membrane module so that the plate is in a position
vertical to the flow direction.
• The filter membrane module and the slide must be completely sub-
mersed (max. 40 cm).
Take changing water levels into account when mounting the filter
membrane module.
• The lower edge of the filter membrane module must be mounted
with least 10 cm distance to the bottom.
Installation
1
Mount the rail of the attachment for filtration in the basin. If nec-
essary, extend the rail with the
M-EXT 1.5 extension accessory.
Fig. 3-14: Sample filtration device (installed)
1 Chain (scope of delivery: Attachment for filtration M 1.5)
2 Guide rail (scope of delivery: Attachment for filtration M 1.5)
3 Height adjustable slide (scope of delivery: Suction line)
4 Suction line (scope of delivery: Suction line)
5 Sleeve tube (scope of delivery: Suction line)
6 Filter membrane module FM with membrane insert
2
Mount the filter membrane module on the slide of the attachment
for filtration.
3
Connect the suction line to the filter membrane module.
In special cases (e.g. in a channel) it is better to mount the
FM filter membrane module in a position horizontal to the
flow direction.
An adapter for horizontal mounting is available as an
accessory.
2
1
3
6
5
4