YSI P 700 IQ Analyzer User Manual
Page 60
60
ba76133e01
09/2013
P 700 IQ
11
Calibrate with calibration standard 1.
11.1 Connect calibration standard 1.
11.2 Select calibration standard 1.
11.3 Start the calibration procedure with
12
Connect calibration standard 2 (see chapter 3.3.5.3 Connecting
the liquid containers, 42).
13
Calibrate with calibration standard 2.
13.1 Connect calibration standard 2.
13.2 Select calibration standard 2.
13.3 Continue the calibration procedure with
The calibration result is displayed after calibration standard 2 has been
measured (Calibration successful!).
Restarting the
measuring operation
14
If necessary, reconnect the calibration standard for automatic
calibration.
15
Carry out the Start measuring function.
Measurement is started and the measured value is displayed in
the measured value display after approx. 5…7 minutes.
16
Carry out the Exit SERVICE function.
or
Exit the calibration and service menu with
17
Confirm the note with
The measured value display appears. The measured value
flashes.
18
If necessary, disconnect the terminal from the MIQ/WCA 232.
19
Close the enclosure of the P 700 IQ.
You have the following possibilities if the calibration failed:
• Repeat the calibration procedure
(make sure that the correct calibration standard is con-
nected and selected for calibration).
• Use the last valid calibration
(see section 4.2.5 Reactivating the last valid calibration,
• Use the factory calibration
(see section 4.2.5 Reactivating the last valid calibration,