4. Measuring / Operation
4.1 Measuring operation
4.1.1 Determination of measured values
The P 700 IQ determines the measured values with a chemical analyz-
ing procedure. Each measurement takes place in several steps.
1
Draw sample and rinse the measuring circle with sample
2
Measure the absorbance of the sample without reagent (blank
value)
3
Add the reagent
4
Leave to react for the reaction time
5
Measure the absorbance of the sample with reagent
The measured values are determined at intervals specified in the set-
tings (see section 4.1.4, 51).
If the intervals overlap with the specified start times and intervals of
other functions (Autom.cleaning or Autom.calibration), the functions
are carried out according to their priority (see section 4.1.4, 51).
4.1.2 Starting the measuring operation
Prior to starting the measuring operation, you always have to check the
readiness for operation of the individual components of the P 700 IQ.
The analyzer does not automatically start measuring when being con-
nected to the IQ S
ENSOR
N
ET
.
Start the measuring operation manually in the calibration and service
menu.
1
Using <>, select the P 700 IQ in the measured value display.
2
Using
The maintenance condition is activated.
3
Confirm the maintenance condition with
The calibration and service menu is displayed.