YSI P 700 IQ Analyzer User Manual
Page 53
ba76133e01
09/2013
53
P 700 IQ
Tab. 4-2: Setting menu
Autom.cleaning
On
Off
Switches the automatic cleaning function
on or off
• Interval
6, 12, 24, 48, 96
Interval in h
• Ref.time hours
0…7…23
Defines the start time from which the
automatic cleaning procedures will take
place at the set interval
(default setting: 7:00 o'clock)
• Ref.time minutes
0…59
Autom.calibration
On
Off
Switches the automatic calibration func-
tion on or off
• Interval
6, 12, 24, 48, 96
Interval in h
• Ref.time hours
0…8…23
Defines the start time from which the
automatic calibration procedures will
take place at the set interval
(default setting: 8:00 o'clock)
• Ref.time minutes
0…59
Status (P700IQ-PO4)
A log book message with the current
status is output.
The status indicates the current situation
of the P700IQ-PO4, e.g.
• Cleaning (ON/OFF)
• Calibrating (ON/OFF/WAIT CLEAN)
• Measuring (ON/OFF/WAIT
CLEAN/WAIT CAL)
• Filling levels of the liquid containers in
%
Meaning
• ON: Function active
• WAIT: Waiting for a cleaning cycle
(WAIT CLEAN) or calibration cycle
(WAIT CAL) to be performed at a
specified point of time.
Autostart
On
Off
Activate or switch off the automatic start
of the P 700 IQ after a power failure
Save and quit
The P 700 IQ stores all changed settings
and the display switches to the next
higher level.
Quit
The display switches to the next higher
level without storing the new settings.
Setting menu
Possible values
Description