4 electrical data, 5 consumption data, Electrical data – YSI P 700 IQ Analyzer User Manual
Page 104: Consumption data, 4 electrical data 7.5 consumption data

104
ba76133e01
09/2013
P 700 IQ
7.4 Electrical data
7.5 Consumption data
The consumption of chemicals depends on the specified intervals and
on the selected measuring range.
Typical consumption
values
Input voltage
Depending on the selected variant:
115 V AC, 60 Hz
Fuse protection of the
building
16 A with ground fault circuit inter-
rupter
Power consumption
35 - 1100 W depending on the option
• Basic consumption
35 W
• Filtration pump
40 W
• Climate kit
200 W
• Heat tracing of the
intake tube
20 W/m (max. 20 m allowed)
• Heat tracing of the
return tube
20 W/m (max. 20 m allowed)
Power supply
Nominal voltage:
115 V AC ± 10 % / 60 Hz
Power supply:
3-pin, N/L /PE
Cable cross-section of power supply:
Europe: 1.5 ... 4.0 mm
2
USA: AWG 14 ... 12
Fuse protection on the operator side:
16 A maximum
Solution
Sufficient for
Reagent solution RE 2.5
4 months (measuring range A, mea-
suring interval 5 minutes)
4 months (measuring range B, mea-
suring interval 10 minutes)
Cleaning solution CL 1.0
4 months (daily cleaning)
Standard solution ST
4 months (daily calibration)