YSI P 700 IQ Analyzer User Manual
Page 106

106
ba76133e01
09/2013
P 700 IQ
EAM3Bx
Meas. range exceeded or undercut
* Check process
EC13Bx
Automatic calibration error: Calibration standard could not be deter-
mined or is not suitable for current measurement range. System
stopped!
* Check the current measurement range
* Check the used calibration standard
EC13Bx
Sensor could not be calibrated,
Sensor blocked for measurement
* Check calibration conditions and calibration standard
* View calibration history
* Service sensor immediately
(see operating manual)
EI13Bx
Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
EI23Bx
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
ES13Bx
Component hardware defective
* Contact service
IC13Bx
Factory calibration has been activated. Make sure the P700IQ-PO4
operates correctly.
IC13Bx
Sensor has been successfully calibrated
* For calibration data, see calibration history
IC23Bx
Last valid user calibration has been activated. Make sure the
P700IQ-PO4 operates correctly.
IC33Bx
Invalid user calibration has been replaced by last valid user calibra-
tion. Caution! Wrong measured values possible. Carry out a new
successful calibration to make sure the P700IQ-PO4 operates cor-
rectly.
IC43Bx
The invalid user calibration was replaced by the factory calibration.
Caution! Wrong measured values possible. Check whether the
P700IQ-PO4 operates correctly or carry out a new successful user
calibration.
Message code
Message text