4 adjustment with signal conditioning instrument – VEGA VEGAPULS 81 User Manual
Page 67
VEGAPULS 64 and 81
67
6.4 Adjustment with signal
conditioning instrument
In general the radar sensors are adjusted with
the PC and the adjustment software VVO.
Radar sensors with digital signal output how-
ever can be also adjusted with the adjustment
and indication module of the signal condition-
ing instruments VEGAMET 514 V or 515 V,
however not the sensors with 0 … 20 mA-
signal output (compact instruments).
Set-up
Sensors with digital output signal are con-
nected with a two-wire cable to a signal condi-
tioning instrument VEGAMET 514 V or 515 V or
to the processing system VEGALOG 571. Via
this two-wire line also the communication with
the sensor is transmitted beside the power
supply and the measuring signal.
The adjust with the adjustment module the
signal conditioning instrument and the con-
nected radar sensors.
Indication and adjustment module of a signal conditioning instrument VEGAMET
513 … 515
ESC
OK
–
+
1
2
on
100
%
CONNECT
515 V
Acc. to the parameter, change value or
select out of list
Select menu point parameter or shift
flashing cursor
Acc. to the menu point, save the
adjusted value or go to the lower menu
Branch, i.e. jump into the lower
menu with [OK]
Acc. to the menu point, cancel
adjustment or jump into the
upper menu
Display,
indication of
- measured value
- menu point
- parameter
- value
Analogue LED-indication
(0 … 100 %)