Types and versions: Configuration of measuring systems

4

4

4

VEGAMET

514V

VEGATRENN

547

VEGAMET

515V

VEGATRENN

547

4

4

Measuring system with VEGAPULS 81 sensors (type A); one sensor
per four-wire line via separator VEGATRENN 547 on the signal con-
ditioning instrument VEGAMET 514 V or 515 V

• Digital output signal; signal processing in the signal conditioning instrument.
• One sensor on one four-wire line. The four-wire line transmits the power supply, the digital

output signal as well as the superimposed adjustment signal.

• Approved in Ex Zone 0. Supply/output signal in EEx i (intrinsically safe).

• Adjustment and parameter adjustment via a PC and the adjustment program VVO- The PC is

connected to the signal conditioning instrument via the adapter VEGACONNECT. In addition
adjustment and parameter adjustment is possible with the signal conditioning instrument.

• An indicating instrument with digital and analogue indication is optionally integrated in the

sensor.

• Optionally an external digital/analogue indication. The indication can be separated from the

sensor in Ex zone 0 up to 25 m.

• The permissible max. resistor of the four-wire line is 7,5

per wire or max. 1000 m cable

length between sensor and safety barrier/signal conditioning instrument.

VEGADIS 10

Signal conditioning

instrument VEGAMET
514 V with separator

VEGATRENN 547 in

housing type 506

VEGACONNECT

PC with adjustment

software VEGA Visual
Operating

VEGAPULS 81

with digital
signal output

VEGADIS 10

VEGAPULS 81
with digital

signal output

VEGACONNECT

PC with adjustment

software VEGA Visual
Operating

Signal conditioning
instrument VEGAMET

515 V with separator
VEGATRENN 547 in

housing type 506

Screened sensor cable for strong
electromagnetic interferences

Current outputs
Voltage outputs
Relay outputs
Fail safe relay
Digital connectability

Outputs (see also Product

Information “signal condi-
tioning instruments series

500“)

Current outputs
Voltage outputs
Relay outputs
Fail safe relay
Digital connectability

Outputs (see also Product

Information “signal condition-
ing instruments series 500“)

Screened sensor cable for strong

electromagnetic interferences

Ex-area

Non-Ex-area

EEx i

EEx i

EEx i

EEx i

EEx i

EEx i

EEx i