Mounting and installation: Installation fault

Pipe antenna on a surge pipe without
ventilation hole

Pipe antenna systems must be provided with a

breathing hole on the upper edge of the surge
pipe. A missing hole will cause faulty measure-

ments.

Pipe antenna: The surge pipe oen to the bot-

tom must have a ventilation or breathing hole

on top

VEGAPULS 64 on a surge pipe: The polariza-
tion direction is in line with the type plate. The

sensor must be directed with the type plate to

the rows of holes

Wrong

Wrong mounting on the vessel with parabolic

tank ceiling

Pipe antenna, wrong polarization
direction

When measuring in a surge pipe, especially if

there are holes in the pipe for mixing, it is

important that the radar sensor is directed to

the rwo of holes.

The two rows of holes of the surge pipe dis-

placed by 180

°

must be in line with the polari-

zation direction of the radar signals. The po-

larization direction is either in line with position
where the type plate is fixed (VEGAPULS 64)

or in line with the casting nose on the lower

part of the housing (VEGAPULS 81).

Type plate

Correct

Wrong

Correct

Wrong