Ohaus TP Series Electronic Balance User Manual
Page 103
50
RS232 COMMAND TABLE
RS232 COMMAND TABLE
RS232 COMMAND TABLE
RS232 COMMAND TABLE
RS232 COMMAND TABLE
Command
Command
Command
Command
Command
Character
Character
Character
Character
Character
Description
Description
Description
Description
Description
?
?
?
?
?
Print current mode
nnnA
nnnA
nnnA
nnnA
nnnA
Set Auto Print feature to “nnn”
(see table).
C
C
C
C
C
Begin span calibration
xD
xD
xD
xD
xD
Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON)
E
E
E
E
E
Exit parts counting or percent weighing
x
x
x
x
x
IIIII
Set Averaging Level to “x”,
where x = 0 to 3 (see table).
L
L
L
L
L
Begin linearity calibration
M
M
M
M
M
Same effect as pressing mode button
xM
xM
xM
xM
xM
Places balance in mode “x”,
where x = 1 to 13 (see table).
If unit or mode is not already
enabled, command will be ignored.
Field: Mode Stab CR
LF
Length:
5
1
1
1
blank if stable
“ ? ” if unstable
1
=
grams
2
=
pennyweight
3
=
carats
4
=
avoidupois ounces
5
=
troy ounces
6
=
grains
7
=
taels
8
=
momme
9
=
pounds
10
=
pounds:ounces
11
=
custom unit
12
=
parts counting
13
=
percent weighing
nnn = 0
Turns feature OFF
nnn = S
Output on stability
nnn = C
Output is continuous
nnn = 1-256
Sets Auto Print
Interval
0
=
minimum level
1
=
2
=
3
=
maximum level
Grams
Pennyweight
Carats
Avoidupois ounces
Troy ounces
Grains
Taels
Momme
Pounds
Pounds:ounces
Custom unit
Parts counting
Percent weighing
Error