Ohaus RANGER ADVANCED COMPACT COUNTING SCALES Manual en User Manual
Page 56
Ranger RP
56
Event and error messages
•
Unstable weight value when
referencing
Ensure stable surroundings
Ensure that the weighing pan is
freely movable
Adjust vibration adapter
•
Impermissible target value or
impermissible tolerance
Repeat input with permissible
values
•
Setting the average piece weight
impermissible during weight
accumulating
End weight accumulating
Reset average piece weight
•
Switching the weighing unit
impermissible during weight
accumulating
End weight accumulating
Switch weighing unit
•
Printout not yet ended
End printout
Repeat required action
•
Switching the weighing unit
impermissible during dynamic
weighing
End dynamic weighing
Switch weighing unit
•
EAROM checksum error
Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
Call OHAUS Service
Weight display unstable
•
Restless installation location
Adjust vibration adapter
•
Draft
Avoid drafts
•
Restless weighing sample
Dynamic weighing
•
Contact between weighing pan
and/or weighing sample and
surroundings
Remedy contact
•
Mains fault
Check mains
Incorrect weight display
•
Incorrect zeroing
Unload scale, set to zero and
repeat weighing operation
•
Incorrect tare value
Clear tare
•
Contact between weighing pan
and/or weighing sample and
surroundings
Remedy contact
•
Scale tilted
Level scale
Error
Cause
Remedy