Settings in the menu

4.3.8

FACT – automatic temperature-dependent adjustment
This menu item appears only on scales with an internal calibration weight.

4.3.9

MIN.WEIG – minimum weight
This menu item appears only if the service technician has saved a minimum weight.

PrOCESS

Adaptation to the weighing process

UNIVEr

Universal setting for all weighing samples and normal weighing goods

dOSING

Dispensing liquid or powdery weighing samples

StAbILI

Adjusting the weighing speed

FASt

The scale operates very fast.

StAndrd

The scale operates at medium speed.

PrECISE

The scale operates with the greatest possible reproducibility.

The slower the scale works, the greater the reproducibility of the weighing results.

TEMP

Defining the temperature difference for automatic calibration

OFF

Switching off automatic calibration in the case of a temperature difference

1K/2K/3K/5K

Automatic calibration in the case of a temperature change of 1 K, 2 K, 3 K or 5 K
since the last adjustment

dAY.tim

Defining up to 7 days of the week and up to 3 times for automatic adjustment.

dAY

Select day of the week for the adjustment.

7 zeros appear in the display after pressing the key

Yes

. The first zero stands for

Monday, the second for Tuesday, the third for Wednesday etc.

Use the key

No

to go to the desired

day of the week and enter 1.

The display 0100100 means that Tuesday and Friday are selected as calibration
days.

Press

Yes

.

tiME1 appears in the display.

tiME1...3

Enter the time(s) for the calibration (hours, minutes).

Note

The format for entering the time (EU or US) depends on the settings in the menu item
TERMINAL-> Device, see section 4.5.1.

ON/OFF

Switching minimum weight function on/off
If the weight on the scale falls below the stored minimum weight, an * appears on
the display in front of the weight indicator.