6 Event and error messages
Error
Cause
Remedy
Display Dark
•
Back lighting set too dark
Set back lighting (b.LIGHt)
brighter
•
No mains voltage
Check mains
•
Unit switched off
Switch on unit
•
Mains cable not plugged in
Plug in mains plug
•
Brief fault
Switch device off and back on
again
Insufficient load
•
Load plate not on the scale
Place load plate on the scale
Set to zero
•
Weighing range not reached
Overload
•
Weighing range exceeded
Unload scale
Reduce preload
•
Result not yet stable
If necessary adjust vibration
adapter or weigh dynamically
•
Function not permissible
Unload scale and set to zero
•
Zeroing not possible with over-
load or insufficient load
Unload scale
•
Average piece weight too low
Select and place larger number of
reference parts on the scale
•
No valid value from the reference
scale
Check cable connection between
the units
Check interface settings
•
No calibration
Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
Calibrate scale
Call OHAUS Service
•
Average piece weight too low
Counting is not possible on this
scale with this average piece
weight