beautypg.com

6 event and error messages, Event and error messages – Ohaus RANGER ADVANCED COMPACT COUNTING SCALES Manual en User Manual

Page 55

background image

Event and error messages

Ranger RP

55

6 Event and error messages

Error

Cause

Remedy

Display Dark

Back lighting set too dark

Set back lighting (b.LIGHt)
brighter

No mains voltage

Check mains

Unit switched off

Switch on unit

Mains cable not plugged in

Plug in mains plug

Brief fault

Switch device off and back on
again

Insufficient load

Load plate not on the scale

Place load plate on the scale
Set to zero

Weighing range not reached

Overload

Weighing range exceeded

Unload scale
Reduce preload

Result not yet stable

If necessary adjust vibration
adapter or weigh dynamically

Function not permissible

Unload scale and set to zero

Zeroing not possible with over-
load or insufficient load

Unload scale

Average piece weight too low

Select and place larger number of
reference parts on the scale

No valid value from the reference
scale

Check cable connection between
the units
Check interface settings

No calibration

Unplug the mains plug then plug
it back in; switch unit off and then
back on in battery mode
Calibrate scale
Call OHAUS Service

Average piece weight too low

Counting is not possible on this
scale with this average piece
weight