3 scale settings (scale), 1 scale1/scale2 - selecting scale, 2 cal - calibration (adjustment) – Ohaus RANGER ADVANCED COMPACT COUNTING SCALES Manual en User Manual
Page 34: Scale settings (scale)

Ranger RP
34
Settings in the menu
4.3 Scale settings (SCALE)
4.3.1
SCALE1/SCALE2 – Selecting scale
This menu item only appears if an analog second scale or a weighing platform is con-
nected.
4.3.2
CAL – calibration (adjustment)
This menu item is not available for certified scales without internal calibration weight.
dIAGNOS
tESt SC
intErN/ExtErN
KboArd
dISPLAY
SNr
SNr2
LiSt
LiSt2
LiSt.M
WOrK.tim
timE
SHOW.tIM
WEIGH
SHOW.WGH
rESEt.AL
SUrE?
Level 1
Level 2
Level 3
Level 4
Level 5
Level 6
Page
Internal
For scales
with an internal calibration weight:
1. Unload scale.
2. Activate menu item CAL with
Yes
. The scale calibrates with the internal calibra-
tion weight. -Int CAL- appears in the display. After calibration is completed,
-donE- appears briefly in the display, and the scale automatically returns to
weighing mode.
External
For scales
without an internal calibration weight:
1. Unload scale.
2. Activate menu item CAL with
Yes
. The scale determines the zero point.
–0- appears in the display. The calibration weight to be placed on the scale then
flashes in the display.
3. If necessary, change the weight value displayed with
No
.
4. Place the calibration weight on the scale and confirm with
Yes
.
The scale calibrates with the calibration weight loaded. After calibration is com-
pleted, -donE- appears briefly in the display, and the scale automatically returns
to weighing mode.