3 counting, 1 counting parts into a container, Counting – Ohaus RANGER ADVANCED COMPACT COUNTING SCALES Manual en User Manual
Page 22: Counting parts into a container

Ranger RP
22
Counting
3 Counting
The compact scales Ranger RP have
additional functions for piece counting.
The relevant settings in the menu are described in Section 4.4.1.
3.1 Counting parts into a container
1. Place the empty container on the scale and press
TARE.
The container is tared and the zero display appears.
2. Place
10 reference parts on the scale and press
Sample 10.
-or-
Place the number of pieces displayed abo
ve the key
Sample Size on the scale
and press
Sample Size
The scale determines the average piece weight and then shows the number of
pieces.
3. Add more parts to the container until the required number of pieces is reached.
4. When the piece counting is completed, press the key
C
to clear the result.
The scale is ready for the next weighing or counting.
Note
•
The average piece weight remains saved in the factory setting until a new average
piece weight is determined.
•
With
Count Weight
it is possible to switch between
the number of pieces and the
weighing units preset.
•
Depending on the assignment, it is possible to display the average piece weight,
i. e. the weight of an individual reference unit, with
Info.
•
If A.CL-APW ON is set in the menu, the average piece weight is automatically
cleared after each counting operation. The average piece weight must be deter-
mined again for the next counting operation.
•
If ACCurCY ON is set in the menu, the accuracy achieved is briefly shown after
the number of pieces is determined.