Technical data, Applications – M&C TechGroup SP2020-H_C_I_BB_F Operator's manual User Manual
Page 8
8
Gas sampling and gas conditioning technology
2-1.1.6.1-ME
7
TECHNICAL DATA
Mains supply
115V 60Hz 800W
Flange
SS316Ti, 3” 150 lbs
Electrical connection
terminals; max 4mm², 2x PG13,5 cable gland
Protective type
IP54 (EN 60529)*
Ambient temperature
thermostat:
-20°C to 60°C
PT100, thermocouple:
-20°C to 80°C
Operating temperature
SP2020-H... : 0-180°C
SP2020-H220... : 0-220°C
Ready for operation
after 2 hours
Temperature status alarm point
30°C to T
Set
*
Alarm contact rating
250V 3A AC, 0,25A DC
Connection sample outlet
1/4"-NPT i., for max. 10mm tube connectors
Connection backflush
tube fitting
3/8” o.d.
Connection calibration gas
tube fitting
¼” o.d.
Filter chamber volume
120 ml
Sample pressure
0,4 to 6 bar abs.
Material of sample conducting parts
SS316Ti, Viton*
alternative = SS316Ti, PTFE
* = Standard
Differential pressure and T
90
-time at different flow rates:
P and T
90
at flow of
100
200
500
1000
1500
Nl/hr
P with new filter element
S-2K 150 or F-0,1GF150:
0,007
0,011
0,020
0,035
0,04
bar
T
90
-time without sample tube/pre filter:
6,0
3,5
1,0
<0,5
<0,5
sec.
8
APPLICATIONS
The SP2020... gas sample probes are used for continuous extraction of gases from dust loaded,
high-temperature or humid processes.
A host of different variations in sampling technology as well as some built-in accessories offer a wide
range of possible applications for nearly all problems encountered in taking gas samples from proc-
esses with high dust content where an effective backflush is necessary. Sampling from processes up
to a maximum of 1900°C are possible with this probe.