Gas sampling and gas conditioning technology

2-1.1.6.1-ME

N O T E !

Attention must be paid to the requirements of IEC 364 (DIN VDE
0100) when setting high-power electrical units with nominal volt-
ages of up to 1000 V, together with the associated standards and
stipulations.

In any case we recommend the use of temperature resistant cable !

A main switch and matching fuse must be provided externally!

The main circuit must be equipped with a fuse corresponding to
the nominal current (over current protection); for electrical details
see technical data.


13.1

STANDARD VERSION WITH INTERNAL CAPILLARY TUBE THERMOSTAT

Remove the lid of the connection box. The electrical connection layout is located in the lid.

Insert the mains cable (min. 3 x 1.5 mm²) through the treaded cable gland and connect to the

appropriate terminals.

Insert the signal cable through the treaded cable connection and connect to the appropriate

terminals.

Screw lid back on.

Figure 5 Electrical connection at the capillary tube thermostat