Gas sampling and gas conditioning technology
2-1.1.6.1-ME
Depending on the process gas temperature and composition, probe tubes made of different
materials are used with ¾" connections:
Material
Type
Process temperature
max. length
Pipe (nipple) o.d.
Stainless steel 316
SS
up to max. 600 °C
2,5 m
25 (37)mm ø
Titan
Ti
up to max. 400 °C
2,5 m *
25 (37)mm ø
Hastelloy
Ha
up to max. 900 °C
2,5 m *
25 (37)mm ø
Incolloy 956
In
up to max. 1200 °C
2,0 m *
25 (37)mm ø
Kanthal
Ka
up to max. 1300 °C
2,5 m *
25 (37)mm ø
Alu. oxide
AO
up to max. 1800 °C
1,5 m *
25 (55)mm ø
PVDF
PV
up to max. 90 °C
1,5 m *
25 (37)mm ø
PTFE
T
up to max. 160 °C
0,5 m
33 mm ø
1m standard *
Depending on the process gas temperature and composition or critical set up situations,
heated probe elements made of stainless steel with flange connection DN65 PN6 are used:
Type
Process temperature
max. length Pipe o.d.
SP30H1, heating max. 320°C
up to max. 550 °C
2,0 m *
40mm ø
SP30H2, heating max. 200°C
up to max. 200 °C
2,0 m *
40mm ø
SP35H, heating max. 320°C
up to max. 550 °C
0,175 m
40mm ø
1m Standard *
In case of higher dust loadings of the process gas, by all means we recommend using a prefilter in
order to prolong service life. This can be supplied with or without dead volume displacer depending
on the response times required. These filters can be screwed directly into the probe flange or via ex-
tension tubes fitted with volume displacers (see Fig.10).
We recommend the following prefilters with ¾" connection:
Dust loading
Prefilter Type
Process Temperature Prefilter o.d.
< 2 g/m³
no filter, only probe tube
see above
2-10 g/m³
/V20, /V20-0
up to max. 600°C
ø 46mm
/V20-3, /V20-4
up to max. 600°C
ø 31mm
/V20/HC, /V20-0/HC
up to max. 900°C
ø 46mm
/V20-5
up to max. 500°C
ø 50mm
with deflector for high flow
speed
60mm ø
> 10 g/m³
/V20-1, /V20-2
up to max. 600°C
ø 60mm
/V20-1/HC, /V20-2/HC
up to max. 900°C
ø 60mm
/V20-T
up to max. 200°C
ø 65mm
/V20-6
up to max. 500°C
ø 60mm
with deflector for high flow
speed
65mm ø
/V12-1
up to max. 600°C
ø 40mm
/V12-2, /V12-3
up to max. 600°C
ø 60mm
/V12-1/SS
up to max. 600°C
ø 40mm
/V12-2/SS, /V12-3/SS
up to max. 600°C
ø 60mm
/V12-1/IC
up to max. 1000°C
ø 40mm
/V12-2/IC, /V12-3/IC
up to max. 1000°C
ø 60mm