The M&C probes SP2600-H.. are designed for stationary use and provide a long service life and a
minimum of maintenance work under the premise of professional selection of the sampling point and
professional mounting.

12.1 MOUNTING OF THE PREFILTER RESPECTIVELY THE SAMPLE TUBE

The pre-filter or sample tube is mounted together with a suitable sealing by screwing into the
G ¾“ thread of the probe flange.

Figure 3

Mounting of the pre-filter or sample tube


12.2

MOUNTING OF THE PROBE

Remove the protection cover of the probe after having opened both bent-level closures.

Put the flange sealing on the bleeder connection.

Fit the mounting piece and the probe flange by means of the attached screws and screw nuts.


In case the bleeder connection does not fit with the standard flange connection DN65 PN6, please
mount the attached optional adapter flange in the same way onto the probe.

N O T E !

It is recommended to mount the probe with its sample gas outlet showing
downwards (not necessary for perfect function).
Further it is recommended to mount the probe with a slight descending
gradient downwards so that possible deposited drops may flow back into the
process.

Remove the heat conducting jaws on the sample gas outlet after having unscrewed the knurled
screw. For connection of the sample line, you have to screw in a correctly dimensioned tube joint
connection 1/4"-NPT by using a PTFE sealing tape.

C A R E !

The fittings must be tightened carefully in order to protect the integrated
components against damage. Do not screw down the fittings to much.
In case of leakage, do not tighten the fittings further. Dismount the
respective fitting completely, screw in and tighten it again.
Check the connection for tightness.