RECEIPT OF GOODS AND STORAGE
Please take the probe and possible special accessories carefully out of the packaging immedi-
ately after receipt and compare the goods with the items listed on the packing list.
Check the goods for any damage caused during delivery and, if necessary, notify your trans-
port insurance company without delay of any damage discovered.
N O T E !
The probe must be stored in a weather-protected and frost-proof area !
12
INSTALLATION AND DIMENSIONS
During installation, the prescriptions for accident prevention and safety instructions for mounting and
operation have to be heeded.
Please strictly observe the notifications of chapter 3 „Safety Instructions“.
Furthermore, you have to consider the following:
Select the optimum sampling point according to the prescriptions actually valid and coordinate
with the responsible persons.
Place the sampling point in such a way that sufficient space for mounting and dismounting of the
probe is available. Also consider the insertion length of the sample tube.
Take care of easy access to the probe in order to enable you to execute any maintenance work
necessary in future without problem.
The bleeder connection must be prepared so that the temperature of the connection piece re-
mains above the acid dew point in order to avoid problems due to corrosion and obstruction.
In case the temperature in the area of the connection piece is >60°C due to radiant heat, you
have to mount a device of sheet steel in order to reflect the radiant heat.
The mounting flange of the connection pi
ece should be connected with size 3” 150lbs. Should you
desire other dimensions, we can provide you suitable adapter intermediate flanges as option. The
minimum flange size or connection piece diameter respectively is determined by the sample tube
diameter or pre-filter diameter you apply.
We recommend to mount the probe horizontally with an angle of inclination of 10° to the process.
N O T E !
Before mounting the probe, you have to check its suitability on the basis of
the given operating parameters (see type plate).