Gas sampling and gas conditioning technology

4-2.1.2-ME

Sample gas IN

Bulkhead union

Back ferrule 6mm Øi

Back ferrule

10mm Øi

Sealing ring

10mm Øi

Sealing ring

6mm Øi

Insert

4mm Øo

Union nut 10mm Øi

Heated sample line

Adapter

PSS...

C o n n e c t i n g t h e h e a t e d s a m p l e l i n e w i t h s p e c i a l a d a p t e r










Place the special adapter on the PTFE hose according to the drawing seen above;

Place the insert into the PTFE hose;

Place the PTFE hose over the nipple on the bulkhead union ‘sample gas IN’ and screw the
adapter tight by hand;

Screw the adapter with a wrench (size 14) 1 1/4 rounds; for this fix the bulkhead union with a

wrench (size 15);

Place the 10mm tube of the heated sample line on the adapter and fix it by screwing the union nut

tight by hand;

Screw the union nut with a wrench (size 19) 1 1/4 rounds; for this fix the adapter with a

corresponding wrench;


Now the threaded joint is mounted tightness and can be loosen and fixed at any time.