19

Gas sampling and gas conditioning technology

4-2.1.2-ME

Indication

Problem

Possible Cause

Check/Solution

Middle LED on
cooler is green;










LED of the LA
electronics is
green;

Cooler and
gas pump in
operation;
condensate in
sample gas
lead;

Pre-filter clogged up;


Pump hose faulty;

Peristaltic pump
SR25.1 faulty;

Not sufficient drying of
sample gas;

Optional liquid alarm
sensor:
Sensor has not turned
pump off.

Remove filter from condensate lead;
Pump delivering?
Change filter;
Pump not delivering?
Change pump hose (see manual SR25.1,
3-7.1-ME);
OK?
Check peristaltic pump (see manual SR25.1,
3-7.1-ME);

OK?
Check cooler (see manual ECP20-1, 3-.10ME);




Check the LA electronics and if necessary replace.