M&C TechGroup PSS-10_1 Operator's manual User Manual
Page 10

10
Gas sampling and gas conditioning technology
4-2.1.2-ME
Options
The gas conditioning and sampling system PSS-10/1 consists of standard one gas-measuring outlet
terminal. An additional flow meter (DK800R), with needle valve, can be fitted to this terminal, whereby
the adjustment of the terminals is carried out in accordance with the specified volume flow rate (see
3.).
In order to protect additionally connected analysers against fluid irruption, and increase the
operational safety of the entire system, we recommend that a fluid alarm sensor LA 1S be installed.
In such a case, the preliminary filter FP-2T, which is delivered with all models, will be replaced with
the FP-2T-D preliminary filter by the manufacturer (see appendix). The LA-1.4 electronics is located
on the clamp mounting rail in the upper part of the case. In the event of a fluid irruption, the LA
electronics automatically turns off the sample gas pump. The alarm will be raised by means of a red
LED. If the equipment is functioning properly, i.e. no alarm, a green LED will be on.
The PSS-10/1 can also be equipped with an additional gas measuring inlet terminal (see figure 3,
Article No. 01G9062 and 01G9063) in order to connect a heated sample line. The existing anti-kink
device is only to be used for heated sample lines in conjunction with the model "C" connection (Article
No. 03 B 1012). Mounting instructions can be found in the Appendix.
The heated sample line, Article No. 01 B 4035, can also be fitted in conjunction with the sample gas
probe PSP 4000.
In the event that the heated line is ordered as additional equipment, the necessary temperature
controller 701 (Article No. 01 G 9055(a)) will be fitted to enclosure by the manufacturer.
A 3-way ball valve (01 G 9046) or 5-way ball valve (01 G 9045) can be fitted to the inlet terminal of
the gas conditioning system in order to calibrate analyser(s) with check gas, or to switch from one
sample gas measurement to another.