Manual prime, Disconnect, Starting on insulin 49 – Energizer 715 User Manual
Page 71

Starting on insulin 49
Manual prime
Manual prime fills the infusion set tubing with insulin before you attach it to the infusion site.
Manual prime is only available after you rewind your pump.
1
After you rewind your
pump, the PREPARING
TO PRIME screen will
appear.
WARNING: Make sure the infusion set is disconnected from your body before you press
ACT
to prime the pump. Never insert the reservoir into the pump while the
tubing is connected to your body. Doing so could result in an accidental
infusion of insulin.
2
Press and hold
ACT
to
start the prime. The
pump will beep 6 times
indicating the prime has
started.
3
While you hold
ACT
, the
pump will beep again 6 times
as the screen begins counting
the prime units being used.
ATTENTION: If you release the
ACT
button:
a.
Press
ACT
. The Manual Prime
screen appears.
b.
Press
ACT
again to continue
the prime sequence (screen
shows where you left off), or
Press
ESC
to rewind.
4
Continue to hold
ACT
until
insulin droplets form on the tip
of the infusion set needle, then
release. Be sure no bubbles are
in the tubing.
PREPARING TO PRIME
Disconnect from Body
Insert/Lock Reservoir
Hold ACT to Prime
The pump may automatically return to the HOME
screen while you install the reservoir. Press
ACT
to
return to the PREPARING TO PRIME screen.
ACT
(hold)
PREPARING TO PRIME
DISCONNECT
Hold ACT
ACT
(hold)
PRIMING – HOLD ACT
ESC if
done
U
0.3
ACT
(hold)